UltraScan Version

Manual


SOMO HPLC-SAXS Module Options Module

Last updated: August 2015

SOMO HPLC-SAXS options

The first set of options deals with baseline removal tools. By deafult, the US-SOMO HPLC-SAXS module will utilize an Integral of I(t) baseline removal tool. The method is based upon the assumption that capillary fouling deposits are formed in proportion to the sample intensity while exposed to the beam (and that the buffer is not responsible). We have developed a mathematical model of this condition and implemented it as the integral baseline procedure. This is an iterative procedure and the number of steps can be set in the Maximum iterations: field (default: 5 iterations). As part of this procedure, the alpha value corresponds to the intensity fouling deposits constant and an early termination criteria can be defined here which will stop the iterative procedure when the difference in alpha between iterations is less than or equal to the alpha early termination limit field.
The program will generate a baseline at each iterative step. By selecting the Produce separate baseline curves checkbox, each baseline will be generated and can be shown in the graphics window. This could be helpful to check if the integral baseline tool has reached convergence. However, it should be deselected before applying the integral baseline tool to multiple files, to avoid the unnecessary generation of many baseline files.
For special needs or very minor baseline correction requirements, the original linear baseline implementation (Brookes et al., J. App. Cryst. 46:1823-1833, 2013) is available by selecting the Linear baseline removal checkbox. A description of the linear baseline removal tool can be found here)

The next set of options deals with the Gaussian mode available. Four alternative checkboxes are available:

The Maximum absolute value of EMG and GMG distortions: field sets limits to the distortions to avoid unreasonably skewed Gaussians (default: 50).

The Clear cached Gaussian values button allows to clear all Gaussian data produced during the current session.

In the Miscellaneous options section there is the option of saving the *.csv files in transposed format, by selecting the Save CSV transposed checkbox.

It is followed by the I(t) negative integral check window field, where the sliding window size for the negative integral test can be set (default: 25). As described previously, this is the size of a sliding window over adjacent frames. If the sum of I(t) values within the window is less than the negative of sum of the corresponding SD values, the I(t) curve will be identified and a warning will be issued. Such regions can be indicative of not optimal buffer subtraction, and might cause problems with the integral baseline correction.

The next line contains the checkbox for the On Make I(t), discard I(t) with no signal above std. dev. multiplied by: and its associated field (default: checked with 3 as the multiplier value). This tests whether or not there is any point in the I(t) curve where the I(t) value is greater than the SD of the point of I(t) multiplied by this field. If the test fails, the I(t) curve is assumed to contain no signal and is dropped with an appropriate messagebox.

The Limit Guinier Maximum q*Rg allows to set this limit to a specified value (default: 1.1) when performing the linear regression in the Guinier analysis to calculate the Rg and I(0) from the I(q) vs q datasets in the Test I(q) module (see the corresponding section in the main Help).

Finally, the last three fields allow modifying the constants for the approximate calculation of the molecular weight as described by Rambo and Tainer (Accurate assesment of mass, models and resolution by small-angle scattering. Nature 496:477-481, 2013). The first two fields contain the constants MW[RT]k (default: 1) and MW[RT]c (default: -2.095), which are the linear fit power law constants as described in eq. (2) of the aforementioned reference and whose default values are taken from Fig. 3, where they were reported for I(q) vs. q data limited to q = 0.3 Å-1. The third field contains the actual qmax cut-off value to which the available I(q) vs. q data will be limited (default: 0.2 Å-1). The constants default values are for globular proteins, and the cut-off value was found empirically to provide better results. If any of these values is changed, a warning will pop-up:

Rambo & Tainer constants changed warning

If the values are not reverted to the default values, this pop-up will keep showing up during the current US-SOMO session every time a molecular weight calculation with the Rambo-Tainer approach is performed, unless the Do not show this warning again button is pressed, avoiding the nuisance when using different values for different kind of biomacromolecules (e.g., RNA) or for testing purposes.


www contact: Emre Brookes

This document is part of the UltraScan Software Documentation distribution.
Copyright © notice.

The latest version of this document can always be found at:

http://somo.uthscsa.edu

Last modified on August 21, 2015.